YX-HRD4201 Ash Fusion Tester
Standard instrument configuration:
1 computer + 1 printer + 1 set of YX-HRD4201 ash fusibility tester
Category:
Physical Properties Analysis Series
Determination of Ash Fusion Temperature of Coal
Keywords:
Analysis instrument
- Product Description
- Technical Features
- Technical Specifications
-
- Commodity name: YX-HRD4201 Ash Fusion Tester
- 仪器标准配置: 1 computer + 1 printer + 1 set of YX-HRD4201 ash fusibility tester
- 技术特点: The furnace chamber is custom‑made from imported materials, with a maximum operating temperature of up to 1600°C. It offers excellent thermal insulation, low surface temperatures, and high structural strength, ensuring both superior heat retention and energy efficiency while effectively preventing deformation or damage caused by external impacts.
It is suitable for measuring the ash fusibility of coal, biomass, and other samples in industries such as power generation, coal, metallurgy, petrochemicals, environmental protection, cement, papermaking, geological exploration, and research institutions.
Scope of Application
It is suitable for measuring the ash fusibility of coal, biomass, and other samples in industries such as power generation, coal, metallurgy, petrochemicals, environmental protection, cement, papermaking, geological exploration, and research institutions.
Conform to the standards
GB/T 219 “Method for Determination of Coal Ash Fusion Temperature”
GB/T 30726 “Method for Determining the Ash Fusion Temperature of Solid Biomass Fuels”
Key words: -
◎ The furnace chamber is specially crafted from imported materials, capable of withstanding temperatures up to 1550°C. It offers excellent thermal insulation, low surface temperatures, and high structural strength, ensuring superior heat retention, energy efficiency, and effective protection against deformation or damage caused by external impacts.
◎ Equipped with a 5-megapixel industrial camera, it provides real-time monitoring of the entire testing process and automatically identifies the four characteristic temperatures of the sample.
◎ It employs a proprietary sealing method and can use the carbon‑sealing technique to ensure the test atmosphere.
◎ Utilizes silicon carbide tube heating, with a maximum operating temperature of 1550°C, enabling a broader range of sample types and more accurate results.
◎ Can be integrated with the laboratory management system to enable backup and upload of test results.
-
Number of samples: 1–5 per batch
Test temperature control range: 200°C to 1550°C
Temperature control accuracy: ±1℃
Experimental atmosphere: oxidizing, weakly reducing (carbon‑sealed method)
Precision: Meets the requirements of GB/T 219 and GB/T 30726.
Accuracy: DT ≤ 40, ST, HT, FT ≤ 30
Power supply: AC 220V ± 22V / 50Hz
Power: ≤4 kW
External dimensions (mm): 820 × 460 × 540
Overall machine weight (kg): 65
Technical Parameter Comparison Table for the Physical Property Analysis Series

Previous page
Our service network covers the entire country.
24-hour on-call service to resolve issues promptly.
We have service centers nationwide, offering prompt in-home installation and guidance to resolve after-sales issues.
One-year warranty, lifetime maintenance, and free software upgrades, installation, and commissioning.
24-hour service, free operational training—ensuring customers enjoy five-star, worry-free after-sales support.
Previous page
Related products
Consult now