YX-DL8500 Automatic Sulfur Analyzer

Standard instrument configuration:

1 computer + 1 printer + 1 set of YX-DL8500 host unit

Category:

Elemental Analysis Series

Determination of Total Sulfur in Coal

Keywords:

Analysis instrument

  • Product Description
  • Technical Features
  • Technical Specifications
    • Commodity name: YX-DL8500 Automatic Sulfur Analyzer
    • 仪器标准配置: 1 computer + 1 printer + 1 set of YX-DL8500 host unit
    • 技术特点: It adopts a horizontal integrated design, ingeniously integrating components such as the automatic sample‑feeding and injection mechanism, the high‑temperature pyrolysis furnace, the electrolytic cell, the stirrer, and the air‑purification system, thereby making the instrument’s structure more compact, more rational, and more stable.

    Based on the principle of coulometric titration, this method is suitable for determining the total sulfur content in combustible materials such as coal, coke, petroleum, and solid biomass fuels, as well as the sulfur content in cement.

    Scope of Application

    Based on the principle of coulometric titration, this method is suitable for determining the total sulfur content in combustible materials such as coal, coke, petroleum, and solid biomass fuels, as well as the sulfur content in cement.

     

    Conform to the standards

    GB/T 214 “Method for Determination of Total Sulfur in Coal”

    GB/T 43218 “Guidelines for Performance Acceptance of Coal Sulfur Analyzers”

    JJG 1006 “Verification Procedure for Instruments for Determining Total Sulfur in Coal”

    GB/T 483 “General Provisions for Coal Analysis Test Methods”

    GB/T 387 “Method for Determination of Sulfur Content in Dark Petroleum Products”

    Key words:
  • ◎ Adopting a horizontal integrated design, the instrument seamlessly integrates components such as the automatic sample‑feeding and injection system, the high‑temperature pyrolysis furnace, the electrolytic cell, the stirrer, and the air‑purification system, resulting in a more compact, rational, and stable overall structure.

    ◎ A specially designed flat-bottomed corundum ceramic boat is employed to minimize sample sticking. Program-controlled operation enables automatic heating, temperature regulation, sample introduction, and analysis; results are automatically saved and printed, and samples can be added mid‑run, significantly enhancing work efficiency and enabling unattended operation.

    ◎ Determination is performed using coulometric titration, with the measurement time automatically adjusted according to the sample type; this method offers faster analysis than both the Eschka gravimetric method and the high‑temperature combustion–neutralization method.

    ◎ Employs a linear electrolyte concentration acquisition circuit, effectively addressing the issue of electrolyte over‑consumption.

    ◎ Employs a contactless switch, featuring open-circuit protection, hardware over-temperature protection, and measures to prevent electrolyte backflow.

    ◎ A multi-point dynamic coefficient correction method is employed to ensure higher accuracy in test results for coals with high, medium, and low sulfur contents.

    ◎ Can be integrated with the laboratory management system to ensure that balance data is never stored locally and that test results are backed up and uploaded.

    ◎ It can be used to determine the total sulfur content in samples such as solid hazardous waste.

  • Measurement range: 0.01% to 40%

    Test method: Coulometric titration

    Sulfur measurement resolution: 0.001%

    Single-sample analysis time: approximately 4 minutes.

    Working furnace temperature: 1150°C (coal, coke), 900°C (oil)

    Temperature control accuracy: ±2℃

    Sample mass: 50 ± 5 mg (coal, coke), 100 ± 10 mg (petroleum products)

    Sample delivery method: automatic delivery, sample rejection

    Number of specimens: 19 per batch; specimens may be added or replaced during the test.

    Accuracy: Conforms to GB/T 214 “Method for Determination of Total Sulfur in Coal”

    Power supply: AC 220 V ± 22 V / 50 Hz

    Power: ≤3kW

    External dimensions (mm): 1010 × 500 × 380

    Overall machine weight: 78 kg

     

    Comparison of Main Technical Parameters of Sulfur Analyzers

Our service network covers the entire country.


24-hour on-call service to resolve issues promptly.


We have service centers nationwide, offering prompt in-home installation and guidance to resolve after-sales issues.


One-year warranty, lifetime maintenance, and free software upgrades, installation, and commissioning.


24-hour service, free operational training—ensuring customers enjoy five-star, worry-free after-sales support.